Loading…

Variable Name: DEV_IDX_S2PVRange_Obj_Low_PH

Variable ID: 0x454D

Variable Label: Range Low

Variable Help: If Measurement Type is pH, set Range Low between -2.0 to 16.0 pH If Measurement Type is Redox(ORP), and Sensor model type is Analog, set Range Low between -2000 to 2000 mV If Measurement Type is Redox(ORP) and Sensor model type is not Analog, set Range Low between -1000 to 1000 mV

Variable Handling: R/W

Variable Class: Correction

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|.2f

Variable Edit Format:

Variable Unit:

Variable Max:

Variable Min:

Variable Read Command List

Command Number Command ID Command Name
235 46E7 Read_PH_Configuration

Variable Write Command List

Command Number Command ID Command Name
236 46E8 Write_PH_Configuration