Variable Name: alarm_pv_conf
Variable ID: 0x4010
Variable Label: Output Alarm
Variable Help: Output Alarm- Configures the output alarm. --No Alarm: Do not use the output alarm. --Lower Limit: If the output is below the lower limit for the ouput alarm, the alarm is activated. --Upper Limit: If the output is above the upper limit for the output alarm, the alarm is activated. --Lower/Uppe Limit: If the ouput is either below the lower limit or above the upper limit for the output alarm, the alarm is activated.
Variable Handling: R/W
Variable Class: Device
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | No Alarm |
0x0001 | Lower Limit |
0x0002 | Upper Limit |
0x0003 | Lower/Upper Limit |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
188 | 4093 | read_alarm_config |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
189 | 4094 | write_alarm_config |