Loading…

变量名称: var_ds_eCalibrationMethodEdit

变量ID: 0x41B9

变量标签: Calibration Method

变量描述: Calibration Method 1-POINT LIN Use the 1-point calibration only if full calibration is not possible. In this method, an empty adjustment (level below the monitoring level) is required. Using additional inputs such as product density and measuring path the count rate for the full calibration (level above the monitoring level) is then determined automatically. 2-POINT EXP An empty calibration (level below the monitoring level) and a full calibration (level above the monitoring level) is required for a two-point calibration.

变量处理: 读/写

变量分类: Correction

变量长度: 1

变量数据类型: 枚举

变量枚举值 变量枚举文本
0x0000 1-POINT
0x0001 2-POINT

变量显示格式:

变量编辑格式:

变量单位:

变量最大值: N/A

变量最小值: N/A

变量读取命令列表

命令号 命令ID 命令名称
239 423D cmd_rd_calibration_settings

变量写入命令列表

命令号 命令ID 命令名称
240 423E cmd_wr_calibration_settings