Variable Name: var_ds_eSignalUnlocked
Variable ID: 0x4005
Variable Label: Signal Unlocked
Variable Help: Signal Unlocked You will get alerted when the device is unlocked with a password. Choose WARNING only if the control system is to be alerted when the device was unlocked with the password. OFF The function is switched off. WARNING Once the detector is unlocked with a password, a warning is issued. If "Error Handling" is set to SENSITIVE, then a fault current is output in addition to the alarm.
Variable Handling: R/W
Variable Class: Correction
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | OFF |
0x0001 | WARNING |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
134 | 41EA | cmd_rd_signal_unlocked |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
135 | 41EB | cmd_wr_signal_unlocked |