Variable Name: var_ds_fUpperCalPoint

Variable ID: 0x41C1

Variable Label: Upper Cal Pt -Full

Variable Help: Upper Cal Pt -Full Count rate with full container or at a level above the switching threshold. If you have selected in "Calibration Method" the setting 1-POINT, the value for "Upper Cal Pt-Full" is automatically calculated from the value of "Lower Cal Pt-Empty" and the parameters defined in the "Product Conditions menu". Select "ReadIn" to read-in the count rate, if you have chosen the setting "2-POINT" at Calibration Method. If you already know the value, you can enter it here.

Variable Handling: R/W

Variable Class: Correction

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|.0f

Variable Edit Format: |en|.0f

Variable Unit:

Variable Max:

Variable Min:

Variable Read Command List

Command Number Command ID Command Name
230 4239 cmd_rd_upper_point

Variable Write Command List

Command Number Command ID Command Name
231 423A cmd_wr_calibration_upper_point