Variable Name: var_ds_fUpperCalPoint
Variable ID: 0x41C1
Variable Label: Upper Cal Pt -Full
Variable Help: Upper Cal Pt -Full Count rate with full container or at a level above the switching threshold. If you have selected in "Calibration Method" the setting 1-POINT, the value for "Upper Cal Pt-Full" is automatically calculated from the value of "Lower Cal Pt-Empty" and the parameters defined in the "Product Conditions menu". Select "ReadIn" to read-in the count rate, if you have chosen the setting "2-POINT" at Calibration Method. If you already know the value, you can enter it here.
Variable Handling: R/W
Variable Class: Correction
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|.0f
Variable Edit Format: |en|.0f
Variable Unit:
Variable Max:
Variable Min:
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
230 | 4239 | cmd_rd_upper_point |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
231 | 423A | cmd_wr_calibration_upper_point |