变量名称: var_ds_eSwitchFunctionEdit
变量ID: 0x41CD
变量标签: Switch Function
变量描述: Switch Function You can specify whether the limit switch is used as a High Alarm or as a Low Alarm. The switching direction affects the fail-safe function of the digital output (open collector). To output the alarm as a signal via the open collector, you have to set "Digital Out Function" in the I/O Setup menu to LS ALARM. You have the following setting options: MAX The alarm is triggered when exceeding the switching threshold. MIN The alarm is triggered when falling below the switching threshold MAX + BUILD UP The alarm is triggered if the switching threshold is exceeded (High Alarm). In addition, monitoring is performed which triggers an alarm if wall deposits 1 is detected. Prerequisite for correct function is that both the count rate for empty and the count rate for full was measured. During the measurement of the empty value it must be ensured that no build-up occurs.
变量处理: 读/写
变量分类: Correction
变量长度: 1
变量数据类型: 枚举
变量枚举值 | 变量枚举文本 |
---|---|
0x0000 | MAX |
0x0001 | MAX + BUILD UP |
0x0002 | MIN |
变量显示格式:
变量编辑格式:
变量单位:
变量最大值: N/A
变量最小值: N/A
变量读取命令列表
命令号 | 命令ID | 命令名称 |
---|---|---|
239 | 423D | cmd_rd_calibration_settings |
变量写入命令列表
命令号 | 命令ID | 命令名称 |
---|---|---|
240 | 423E | cmd_wr_calibration_settings |