Variable Name: dvm_fpct
Variable ID: 0x40DC
Variable Label: Pct. A.
Variable Help: The measured percentage of media A in the fluid. Fraction is defined as a part of mixture. Mixtures consisting of two components ( A+ B) which can be measured separately.If the flowmeter is ordered with fraction, e.g °BRIX,the flowmeter is able to calculate the % concentration of sugar in a water (B) + sugar (A) mixture.From factory the table slope 'b' is set to 1.0000. Formula: % concentration = (a / 3600) + bx % concentration = the concentration of media e.g. in °BRIX a = fraction offset in (kg/h) b = is a factor without dimension (table slope) x = is the concentration in % or e.g. °BRIX Example: If you want to change the concentration in % on the flowmeter with +0,5 %, you have to change the table slope 'b' to 1.005. After the change the flowmeter will now show a concentration 0,5 % higher as before.
Variable Handling: R
Variable Class: Correction
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|.1f
Variable Edit Format:
Variable Unit:
Variable Max: 2900
Variable Min: 0
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
134 | 40F4 | read_float32 |
Variable Write Command List
None