Loading…

Variable Name: ao_alarm_out

Variable ID: 0x4027

Variable Label: Analog alarm out

Variable Help:

Variable Handling: R/W

Variable Class: Device

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 2.4mA or less
0x0001 4.0mA
0x0002 Hold
0x0003 21.6mA or more

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
130 4049 read_analog

Variable Write Command List

Command Number Command ID Command Name
167 4066 write_ao_alarm_out