Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_tempAutoTCOption | 402A | Auto TC Option | Determines the algorithm that adjusts for the temperature effect of a solute. | R/W | Enum | 8 |
var_tempAutoTCCoeff | 402C | TC Coeff | Temp comp is based on percent change of conductivity from an adjustable reference temperature per deg C. The value is adjustable between 0 and 9.99 % per deg-C | R/W | Float | 8 |
var_tempCompType | 4028 | Temp.Comp.Type | Determines the type of Temperature Compensation algorithm applied to the Primary Value | R/W | Enum | 8 |
var_manualTemp | 4029 | Manual Temp | Manual Temperature Compensation Value - is used in Nernstian temp. comp. when the Temp Comp selection is Manual | R/W | Float | 8 |
var_refTemp | 402D | Reference Temp | Variable reference temperature - this provides the ability to use any temp as the reference temp used to determine the temp-compensated PV | R/W | Float | 8 |
var_TCUsrDefX1 | 40A7 | Temp 1 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefX2 | 40A8 | Temp 2 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefX3 | 40A9 | Temp 3 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefX4 | 40AA | Temp 4 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefX5 | 40AB | Temp 5 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |