Loading…
Name ID Label Description Type Access Class
var_tempAutoTCOption 402A Auto TC Option Determines the algorithm that adjusts for the temperature effect of a solute. R/W Enum 8
var_tempAutoTCCoeff 402C TC Coeff Temp comp is based on percent change of conductivity from an adjustable reference temperature per deg C. The value is adjustable between 0 and 9.99 % per deg-C R/W Float 8
var_tempCompType 4028 Temp.Comp.Type Determines the type of Temperature Compensation algorithm applied to the Primary Value R/W Enum 8
var_manualTemp 4029 Manual Temp Manual Temperature Compensation Value - is used in Nernstian temp. comp. when the Temp Comp selection is Manual R/W Float 8
var_refTemp 402D Reference Temp Variable reference temperature - this provides the ability to use any temp as the reference temp used to determine the temp-compensated PV R/W Float 8
var_TCUsrDefX1 40A7 Temp 1 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8
var_TCUsrDefX2 40A8 Temp 2 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8
var_TCUsrDefX3 40A9 Temp 3 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8
var_TCUsrDefX4 40AA Temp 4 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8
var_TCUsrDefX5 40AB Temp 5 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8