Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_TCUsrDefX6 | 40AC | Temp 6 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefY1 | 40C3 | K1/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefY2 | 40C4 | K2/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefY3 | 40C5 | K3/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefY4 | 40C6 | K4/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefY5 | 40C7 | K5/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_TCUsrDefY6 | 40C8 | K6/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | R/W | Float | 8 |
var_ConcUsrDefX1 | 40B9 | X1 Cond | Concentration curve with x-axis = Conductivity in µS/cm and y-axis = Concentration in Conc. Unit. | R/W | Float | 8 |
var_ConcUsrDefX2 | 40BA | X2 Cond | Concentration curve with x-axis = Conductivity in µS/cm and y-axis = Concentration in Conc. Unit. | R/W | Float | 8 |
var_ConcUsrDefX3 | 40BB | X3 Cond | Concentration curve with x-axis = Conductivity in µS/cm and y-axis = Concentration in Conc. Unit. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |