Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
PH_S1_transducerSerialNumber | 42AB | PH_S1_transducerSerialNumber | Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. | R/W | Unsigned | 2048 |
PH_S1_maxUpdateTime | 42AC | PH_S1_maxUpdateTime | R | Time | 0 | |
PH_S1_deviceVariableFamilyCodes | 42AD | PH_S1_deviceVariableFamilyCodes | Device Family- Indicates the family (i.e, pressure, temperature, flow, etc.) that the device variable belongs. | R | Enum | 2048 |
PH_S2_Obj_Unit | 42AF | PH_S2_Obj_Unit | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit | R/W | Enum | 0 |
PH_S2_Obj_Value | 42B0 | PH_S2_Obj_Value_LABEL | Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention | R | Float | 10 |
PH_S2_Obj_DQ | 42B1 | PH_S2_Obj_DQ_Label | Process Data Quality- Indicates the overall process data quality of the device variable value. | R | Enum | 10 |
PH_S2_Obj_LS | 42B2 | PH_S2_Obj_LS_Label | Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). | R | Enum | 10 |
PH_S2_Obj_Status | 42B3 | PH_S2_Obj_Status_Label | Device Family Status- Indicates additional status information on a device variable for a specific device family. | R | BitEnum | 10 |
PH_S2_Obj_Classification | 42B4 | PH_S2_Obj_Classification | R | Enum | 0 | |
PH_S2_UpperSensorLim_Obj | 42B5 | usl | Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. | R | Float | 8 |