Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
press_freq_idx_alarm_ena | 40E7 | Press Freq Index Alarm Use | Press Freq Index Alarm Use- Sets whether the Pressure Frequency Alarm is activated, for example when the Pressure Frequency Index is greater than Press Freq High Limit, or when the Pressure Frequency Index is less than Press Freq Low Limit. | R/W | Enum | 1 |
overload_cnt | 40E8 | OOR Pressure Count | OOR Pressure Count- Displays the number of occurrences of out-of-range pressure. This value can be reset. | R | Unsigned | 3 |
overload_cnt_low_limit | 40C9 | Normal Pressure Low Limit | Normal Pressure Low Limit- This is the low limit for normal pressure. After calbration, if the pressure drops below this limit, OOR Pressure Count will increase by 1. | R/W | Float | 1 |
overload_cnt_high_limit | 40CA | Normal Pressure High Limit | Normal Pressure High Limit- This is the high limit for normal pressure. After calbration, if the pressure exceeds this limit, OOR Pressure Count will increase by 1. | R/W | Float | 1 |
overload_cnt_threshold | 40E9 | OOR Count Alarm Threshold | OOR Count Alarm Threshold- Sets the threshold for the Out-of-Range Count Alarm. If OOR Count Alarm Use is set to "Enabled" and the OOR Pressure Count is greater than or equal to this threshold, the alarm will turn ON. | R/W | Unsigned | 1 |
overload_cnt_alarm_ena | 40EA | OOR Count Alarm Use | OOR Count Alarm Use- Sets whether the Out-of-Range Count Alarm is activated, when OOR Pressure Count is greater than or equal to OOR Count Alarm Threshold. | R/W | Enum | 1 |
standard_average_press | 40C8 | Average Pressure | Average Pressure- Displays the average value for the measured pressure. | R | Float | 3 |
standard_deviation | 40CD | Standard Deviation | Standard Deviation- Displays the standard deviation for the pressure. | R | Float | 3 |
standard_deviation_max | 40CE | Standard Deviation Max | Standard Deviation Max- Displays the maximum standard deviation for the pressure in certain period. | R | Float | 3 |
standard_deviation_min | 40CF | Standard Deviation Min | Standard Deviation Min- Displays the minimum standard deviation for the pressure in certain period. | R | Float | 3 |
Name | ID | Label | Description | Type | Access | Class |