Name ID Label Description Type Access Class
production_number 40E2 Production Number Production Number- Displays the pressure transmitter Production Number. R ASCII 0
press_freq_index 40D8 Pressure Frequency Index Pressure Frequency Index- The value indicating Pressure Frequency index. R Float 3
press_freq_index_min 40DA Press Freq Index Min Press Freq Index Min- Displays the minimum value of Pressure Frequency Index during a certain period. R Float 3
press_freq_index_max 40D9 Press Freq Index Max Press Freq Index Max- Displays the maximum value of Pressure Frequency Index during a certain period. R Float 3
press_freq_idx_sensor_select 40D7 Press Freq Index Sensor Selection Press Freq Index Sensor Selection- Displays the sensor (Differential Pressure Sensor or Static Pressure Sensor) used to calculate the Pressure Frequency Index and the sampling period of the sensor if the device is a Differential Pressure Transmitter. Also, displays the sampling period of the Differential Pressure Sensor if the device is a Gauge Pressure Transmitter or Absolute Pressure Transmitter. R/W Enum 1
press_freq_calc_pv_limit_low 40C6 Press Freq Calc PV Low Limit Press Freq Calc PV Low Limit- The Pressure Frequency Index is calculated only when the pressure is greater than or equal to this low limit. R/W Float 1
press_freq_calc_pv_limit_high 40C7 Press Freq Calc PV High Limit Press Freq Calc PV High Limit- The Pressure Frequency Index is calculated only when the pressure is less than or equal to this high limit. R/W Float 1
press_freq_filter_constant 40E4 Press Freq Filter Constant Press Freq Filter Constant- Sets the filter constant used to calculate the Pressure Frequency Index. The larger the index value, the fewer the detection errors, because low-frequency components are more thoroughly removed. However, the amount of variation in the index also becomes smaller, requiring a more strict setting of the alarm. R/W Float 1
press_freq_idx_threshold_low 40E5 Press Freq Index Low Limit Press Freq Index Low Limit- Sets the low limit for the Pressure Frequency Index. If Press Freq Index Alarm Use is set to "Enabled (Low)" or "Enabled (High and Low)," and the Pressure Frequency Index is less than this limit, the alarm will turn ON. R/W Float 1
press_freq_idx_threshold_high 40E6 Press Freq Index High Limit Press Freq Index High Limit- Sets the high limit for the Pressure Frequency Index. If Press Freq Index Alarm Use is set to "Enabled (High)" or "Enabled (High and Low)" and the Pressure Frequency Index is greater than this limit, the alarm will turn ON. R/W Float 1
Name ID Label Description Type Access Class