Loading…
Name ID Label Description Type Access Class
percentRange 00E9 % Meas Range Porcentaje de rango: Variable que rastrea la representación del valor digital con respecto al rango definido por el valor de rango inferior y el valor de rango superior, para modos de funcionamiento normales. Las unidades de esta variable siempre son en porcentaje. R Float 34
transfer_function 1F84 Transfer Function Transfer Function Here you can specify how the current output should behave in case of a change in level: LINEAR A continuous 4-20 mA signal is output which is proportional to the count rate. DISCRETE SWITCH Depending on the set switching threshold (alarm threshold for the level) a current of 4 mA or 20 mA is output. Values between 4 and 20 mA will not be output. R/W Enum 32
upperRange_value 00EA Upper Range Value Valor de rango superior: Define el punto final de funcionamiento del cual se derivan el valor analógico y el punto 100% del rango de porcentaje. Además, el valor de rango superior define un punto final de funcionamiento del que se derivan las alarmas asociadas con el valor analógico y las alarmas asociadas con la representación del valor digital. R Float 32
lowerRange_value 00EB Lower Range Value Valor de rango inferior: Define el punto final de funcionamiento del que se derivan el valor analógico, y el punto 0% del rango de porcentaje. Además, el valor de rango inferior define un punto final de funcionamiento del que se derivan las alarmas asociadas al valor analógico y las alarmas asociadas a la representación del valor digital. R Float 32
response_code 0096 Response code Command Specific Response Code- Data that is returned in the Field Device response, that indicates the relative success of the Command sent in the Host request. This information may relate to the Data of the Host request or to the command itself. R Enum 129
comm_status 0098 Communication status Communications Error Summary- The Field Device has detected the following error condition while attempting to receive the Host request. Therefore the Command was not accepted, and the Host request caused no change to occur in the Field Device. R BitEnum 131
request_preambles 009B Num Req Preams Number of Request Preambles- Number of Preambles required from the Host request by the Field Device R Unsigned 128
universal_revision 009C Universal rev Universal Revision- Revision of the Universal Device Description, that the Field Device conforms to R Unsigned 128
transmitter_revision 009D Field Dev Rev Field Device Revision- Revision of the Field Device Specific Device Description, that the Field Device conforms to R Unsigned 128
software_revision 009E Software Rev Field Device Software Revision- This revision corresponds to the software or firmware, that is embedded in the Field Device R Unsigned 2048
Name ID Label Description Type Access Class