Loading…
名称 ID 标签 描述 类型 访问 分类
var_dd_fProbeCps 4033 Cps Single Detector Cps Single Detector Shows the count rate of the active detector normalized to one second. 浮点 2050
var_dd_fMeasurementCps 4034 Meas CH CPS 浮点 2050
var_dd_fControlCps 4035 Control CH CPS 浮点 2050
var_dd_fAuxiliaryCps 4036 Auxiliary CH CPS 浮点 2050
var_dd_eHVMode 402C HV Mode HV Mode The HV-mode (high-voltage control) provides the following options: AUTO Automatic HV control. Select this mode in normal measurement mode. It enables a temperature-stable operation of the detector. MANUAL Manual HV control. This setting is used for testing purposes; you can also provisionally set the operating point of the detector. The specified voltage must, however, lie in the plateau. If you select "Manual", the automatic HV control is switched off. The HV is then set to the value that was set in the menu item HV Manual. PLATEAU Starts the plateau measurement. If you select PLATEAU, the measurement mode is exited and the plateau of the multiplier measured, which is used in the detector. 读/写 枚举 2050
var_ds_uiPlateauHVStart 4037 HV Start HV Start Enter HV start value for plateau measurement beginning 读/写 无符号 2048
var_ds_uiPlateauHVStop 4038 HV Stop HV Stop Enter HV stop value for plateau measurement end 读/写 无符号 2048
var_ds_uiPlateauHVIncrement 4039 HV Step HV Step Enter HV step width for plateau measurement step width 读/写 无符号 2048
var_ds_uiPlateauMeasuringTime 403A Meas. Time Meas. Time Enter measuring time for each plateau measuring point 读/写 整型 2048
var_ds_uiPlateauTableCount 403B Plateau Tab Entries 无符号 2048
名称 ID 标签 描述 类型 访问 分类