名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
var_dd_fTempMin | 4028 | Device Temp. Min | Device Temp. Min Minimum device temperature since last reset | 读 | 浮点 | 2050 |
var_dd_fTempMax | 4029 | Device Temp. Max | Device Temp. Max Maximum device temperature since last reset | 读 | 浮点 | 2050 |
var_ds_eResetTempExtrema | 402A | Reset Temp Extrema | 读/写 | 枚举 | 8 | |
var_ds_uiDetectorCode | 402B | Detector Code | Detector Code The currently selected detector code is displayed. The detector code is crucial for the proper function of the automatic HV control (high voltage control). The detector code has already been set by BERTHOLD TECHNOLOGIES. It is dependent on the scintillator size and the detector type, see operating manual. | 读/写 | 无符号 | 2048 |
var_ds_fHVManual | 402D | HV Manual | HV Manual You can enter a fixed HV value. The value becomes active when you select MANUAL at "HV-Mode". | 读/写 | 浮点 | 2048 |
var_ds_fHVDefault | 402E | HV Default | HV Default This setting affects: - A quick operation point control after power failure After a power failure, the HV starts at the last "HV Average" value which results from the "HV Default". This will significantly reduce the start-up phase after power failure. - Error detection in case of HV-drift If the HV deviates too far (+40%, -20%) from the HV Default or if the HV reaches the limits of the range of 300 V or 1300 V, this is reported as an error. - Special feature: 0V For maintenance purposes, the HV Default can also be set to 0V. In this case, the mechanisms described above are suspended; after a power failure, the measurement begins to get adjusted at 800 V. "HV Default" has already determined and set by BERTHOLD TECHNOLOGIES. | 读/写 | 浮点 | 2048 |
var_dd_fHVLive | 402F | HV Live | HV Live Displays the current HV (high voltage) at the photomultiplier. If the "HV Mode" is set to AUTO (normal operation), the values in HV Live must change, viewed over several seconds. | 读 | 浮点 | 2050 |
var_dd_fHVLiveMaster | 4030 | HV Live | HV Live Displays the current HV (high voltage) at the photomultiplier. If the "HV Mode" is set to AUTO (normal operation), the values in HV Live must change, viewed over several seconds. | 读 | 浮点 | 2050 |
var_dd_fHVFeedback | 4031 | HV Feedback | 读 | 浮点 | 2050 | |
var_dd_fHVAverage | 4032 | HV Average | HV Average Shows the average HV (high voltage) of the last 10 days. Significant deviations from "HV Live" to "HV Average" are considered and reported as errors. | 读 | 浮点 | 2050 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |