var_dd_fTempMin 4028 Device Temp. Min Device Temp. Min Minimum device temperature since last reset 浮点 2050
var_dd_fTempMax 4029 Device Temp. Max Device Temp. Max Maximum device temperature since last reset 浮点 2050
var_ds_eResetTempExtrema 402A Reset Temp Extrema 读/写 枚举 8
var_ds_uiDetectorCode 402B Detector Code Detector Code The currently selected detector code is displayed. The detector code is crucial for the proper function of the automatic HV control (high voltage control). The detector code has already been set by BERTHOLD TECHNOLOGIES. It is dependent on the scintillator size and the detector type, see operating manual. 读/写 无符号 2048
var_ds_fHVManual 402D HV Manual HV Manual You can enter a fixed HV value. The value becomes active when you select MANUAL at "HV-Mode". 读/写 浮点 2048
var_ds_fHVDefault 402E HV Default HV Default This setting affects: - A quick operation point control after power failure After a power failure, the HV starts at the last "HV Average" value which results from the "HV Default". This will significantly reduce the start-up phase after power failure. - Error detection in case of HV-drift If the HV deviates too far (+40%, -20%) from the HV Default or if the HV reaches the limits of the range of 300 V or 1300 V, this is reported as an error. - Special feature: 0V For maintenance purposes, the HV Default can also be set to 0V. In this case, the mechanisms described above are suspended; after a power failure, the measurement begins to get adjusted at 800 V. "HV Default" has already determined and set by BERTHOLD TECHNOLOGIES. 读/写 浮点 2048
var_dd_fHVLive 402F HV Live HV Live Displays the current HV (high voltage) at the photomultiplier. If the "HV Mode" is set to AUTO (normal operation), the values in HV Live must change, viewed over several seconds. 浮点 2050
var_dd_fHVLiveMaster 4030 HV Live HV Live Displays the current HV (high voltage) at the photomultiplier. If the "HV Mode" is set to AUTO (normal operation), the values in HV Live must change, viewed over several seconds. 浮点 2050
var_dd_fHVFeedback 4031 HV Feedback 浮点 2050
var_dd_fHVAverage 4032 HV Average HV Average Shows the average HV (high voltage) of the last 10 days. Significant deviations from "HV Live" to "HV Average" are considered and reported as errors. 浮点 2050