Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_ds_dtPlateauTableDate | 403C | Plateau Date | R | HartDate | 2048 | |
var_ds_uiPlateauTableIndex | 403D | Plateau Index | Plateau Index Active plateau table Index | R/W | Unsigned | 2048 |
var_ds_fPlateauHV | 403E | Plateau HV | Plateau HV HV of active displayed plateau table entry | R | Float | 2048 |
var_ds_fPlateauImpulses | 403F | Plateau Impulses | Plateau Impulses Impulses of active displayed plateau table entry | R | Float | 2048 |
var_dd_fPMTCurrent | 4040 | PMT Current | PMT Current Displays active current at multiplier | R | Float | 2050 |
var_dd_fPMTCurrentMax | 4041 | PMT Current Max | PMT Current Max Displays the maximum PMT current measured since last reset. | R | Float | 2050 |
var_ds_ePMTReset | 4042 | Reset Current Extrema | R/W | Enum | 8 | |
var_ds_fErrorCurrent | 4043 | Error Current Value | Error Current Value Here you define the fault current in mA, if you have selected Value in the "Loop Alarm Type" parameter. If you have selected High or Low, then the appropriate current value is displayed here. If you exceed the mA value, then Value is set in the "Loop Alarm Type" parameter. | R/W | Float | 64 |
var_ds_fCurrentLowerLimit | 4044 | Current Lower Limit | Current Lower Limit Lower limit of the current range for the 4-20 mA current output. For adjustment purposes or to safely identify an over or under measuring range condition, the current range available for the measurement signal is extended beyond the standard range of 4 mA and 20mA. According to the Namur specifications (NE 43), the lower current value must not be less than 3.8 mA. The limits can be set within the range from 3.8 mA to 4 mA. | R/W | Float | 64 |
var_ds_fCurrentUpperLimit | 4045 | Current Upper Limit | Current Upper Limit Upper limit of the current range for the 4-20 mA current output. For adjustment purposes or to safely identify an over or under measuring range condition, the current range available for the measurement signal is extended beyond the standard range of 4 mA and 20mA. According to Namur specifications (NE 43), the upper current value must not be higher than 20.5mA. The limits can be set within the range from 20 mA to 20.5 mA. | R/W | Float | 64 |
Name | ID | Label | Description | Type | Access | Class |