Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_dd_eDOValue | 4046 | Digital Out State | R | Enum | 1026 | |
var_dd_eDOMode | 4047 | Digital Out Mode | R/W | Enum | 1026 | |
var_ds_eDOTestValue | 4048 | Digital Out Test State | R/W | Enum | 1024 | |
var_dd_eDIValue | 4049 | Digital In State | R | Enum | 1026 | |
var_dd_eDIMode | 404A | Digital In Mode | R/W | Enum | 1026 | |
var_ds_eDITestValue | 404B | Digital In Test State | R/W | Enum | 1024 | |
var_ds_uiResponseDelay | 404C | Meas Delay Time | Meas Delay Time If interference radiation detection is enabled, the measured values are output with a delay. The delay is adjustable from 0 to 5 seconds. The delay is required so that at the moment of detection the measured value remains unaffected, because in this case, increased count rates are already present which may distort the measured value. | R/W | Unsigned | 2048 |
var_ds_fIoFactor | 404E | Io Factor | Io Factor Defines the threshold where the interference radiation detection responds. With the default value of 1.5, the threshold is 1.5 times the active empty count rate and is suitable for most applications. You can adjust the factor for the threshold yourself. Note that a value below 1.5 may trigger faulty switching caused by fluctuations in the measured values. Increasing the factor decreases the sensitivity. Increase the factor if false alarms are triggered by significant fluctuations in the measured values, for example, by agitator blades. | R/W | Float | 2048 |
var_ds_fWaitingTime | 404F | Waiting Time | Waiting Time If interference radiation is detected, the measurement is "frozen" and will be released at the earliest after the waiting period is over. You have to enter the value for the waiting time in seconds. | R/W | Float | 2048 |
var_dd_eCpsAverageMode | 4050 | Cps Average Mode | Cps Average Mode Select cps average mode. Select 'Fix Value' for test. | R/W | Enum | 2050 |
Name | ID | Label | Description | Type | Access | Class |