Name ID Label Description Type Access Class
var_ds_dtSourceExchangeDate 405B Warning Date Warning Date Enter the date when you want to get the Source Exchange warning (format: MM/DD/YYYY) R/W HartDate 2048
var_dd_fReadInRemainedTime 405D Remaining Time R Float 10
var_ds_fReadInTime 405E ReadIn Time ReadIn Time Here you can define the period of time over which the count rate is to be read-in for each calibration point. The statistical variation of the count rate is averaged over this time period. A longer read-in time results in a better mean value. A time period of 30 sec (default setting) is usually reasonable. R/W Float 8
var_ds_eApplicationLevel 405F Application Level R Enum 2048
var_ds_ascSWVersion 4060 SW Revision R ASCII 2048
var_ds_ascSWVersionDate 4061 SW Revision Date R ASCII 2048
var_ds_eLoopMonitorSwitch 4062 Current Loop Monitoring Current Loop Monitoring Enable or disable the monitoring of the 4-20mA current signal. Monitoring checks whether the set current is actually flowing in the current loop, and signals an error if any deviation is detected. You have the following setting options: ENABLED Monitoring is enabled (default setting). Unless there are compelling reasons, you should keep this setting. The setting is automatically enabled when the Safety Mode is selected. DISABLED Monitoring is disabled. The HART Communicator outputs a corresponding message. R/W Enum 2048
var_ls_fPlateauPointHV0 4064 R Float 4096
var_ls_fPlateauPointHV1 4065 R Float 4096
var_ls_fPlateauPointHV2 4066 R Float 4096
Name ID Label Description Type Access Class