Loading…
Name ID Label Description Type Access Class
temperature_value 4007 S.tmp. Sensor Temperature- Temperature of the sensor that is making the process measurement R Float 10
temperature_units 4008 temp unit Sensor Temperature Units- Engineering unit to be displayed with Sensor Temperature. R/W Enum 8
analog_value 4009 I Out Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes R/W Float 66
analog_alarm_code 400A AO Alrm typ Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. R Enum 64
sensor_serial_number 400B PV Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
min_value_messuret 4002 Min P Datalogger function, stores min value measured. Value can be cleared (Reset min/max) R/W Float 10
max_value_messuret 4001 Max P Datalogger function, stores max value measured. Value can be cleared (Reset min/max) R/W Float 10
under_min_sensor_value 400E Under Lim Input under LRV indication 3.50 .. IOUT0% mA R/W Float 64
over_max_sensor_value 400F Over Lim Input over URV indication IOUT100% .. 23.00 mA R/W Float 64
sensor_offset_value 4003 S.off. The offset added to the value measured R/W Float 8
Name ID Label Description Type Access Class