Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
varIDString | 401F | ID String | XgardIQ identification string. | R/W | ASCII | 0 |
varRelayFit | 4050 | Relays Fitted? | Optional fitted component. | R | Enum | 0 |
varFFFit | 4064 | Foundation Fieldbus Fitted? | Optional fitted component. | R | Enum | 0 |
varSensorTech | 4020 | Sensor | Gas sensor module type. | R | Enum | 0 |
varIsCO2 | 408A | R | Enum | 0 | ||
varZSuppress | 402C | Zero Suppression | Level of suppression used to reduced noise around the clean air value. | R/W | Enum | 0 |
varPellCtrl | 4090 | W | Enum | 4 | ||
varBumpRespTimeout | 404A | Gas Sense Time | Time instrument will wait for initial gas response before test start. Valid range 5 to 300 seconds. | R/W | Unsigned | 0 |
varBumpTestTime | 404B | Bump Test Time | Maximum test time for speedy bump; total test time for smart bump. Valid range 5 to 300 seconds. | R/W | Unsigned | 0 |
varBumpLowLimit | 404C | Smart Bump Lower Limit | Lowest gas reading allowed to pass a smart bump. Valid range 0 to sensor range. | R/W | Float | 0 |
Name | ID | Label | Description | Type | Access | Class |