Name ID Label Description Type Access Class
varAoutOperationalInhibit 4038 Inhibit Level mA signal level when in operational inhibit state. R/W Enum 0
varAoutPowerOnInhibit 4039 Power-on Inhibit mA signal level when in power-on inhibit state. R/W Enum 0
varNumCrossGases 4067 Number of Cross Gases Number of gases for cross calibration and bump R Integer 0
varSensorRange 4066 Gas Range Full measurement range of gas sensor. R Float 0
varSensorName 4065 Gas Name Target gas name. R ASCII 0
varSensorUnits 409E Gas Units Target gas measurement units. R ASCII 0
varSensorXLevel_1 406A Cross Gas 1 Level Level of cross gas to apply for cross calibration and bump. R Float 0
varSensorXName_1 4069 Cross Gas 1 Name Gas to be used for cross calibration and bump. R ASCII 0
varSensorXUnits_1 4084 Cross Gas 1 Units Units of gas to be used for cross calibration and bump. R ASCII 0
varSensorXFactor_1 4085 Cross Calbration Factor 1 Relative cross sensitivity to target gas. R Float 0
Name ID Label Description Type Access Class