Loading…
Name ID Label Description Type Access Class
rHighDeadbandAlarmA 40BE High Deadband A Sets the High Alarm deadband for Alarm A. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rHighDeadbandAlarmB 40BF High Deadband B Sets the High Alarm deadband for Alarm B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rHighDeadbandAlarmC 40C0 High Deadband C Sets the High Alarm deadband for Alarm C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rHighDeadbandAlarmD 40C1 High Deadband D Sets the High Alarm deadband for Alarm D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rLowDeadbandAlarmA 40C2 Low Deadband A Sets the Low Alarm deadband for Alarm A. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rLowDeadbandAlarmB 40C3 Low Deadband B Sets the Low Alarm deadband for Alarm B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rLowDeadbandAlarmC 40C4 Low Deadband C Sets the Low Alarm deadband for Alarm C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rLowDeadbandAlarmD 40C5 Low Deadband D Sets the Low Alarm deadband for Alarm D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint R/W Float 0
rSetpointControlA 40C6 Setpoint Control A Sets the Controller Setpoint for Relay A. R/W Float 0
rSetpointControlB 40C7 Setpoint Control B Sets the Controller Setpoint for Relay B. R/W Float 0
Name ID Label Description Type Access Class