Loading…
Name ID Label Description Type Access Class
ePvUnits 4026 PV Units Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 8
eORPUnits 4015 ORP Unit, mV. R Enum 8
rORPValue 4016 ORP The measured value of ORP in mV. R Float 10
rORPUSL 4017 ORP USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
rORPLSL 4018 ORP LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
rORPMinSpan 4019 ORP Min Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
rORPDamping 401A ORP Damping Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R Float 8
epHUnits 4020 pH Unit, pH. R Enum 8
rpHValue 4021 pH The measured value of pH. R Float 10
rpHUSL 4022 pH USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
Name ID Label Description Type Access Class