Name ID Label Description Type Access Class
lower_range_val 401A LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 64
upper_range_val 401B URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 64
man_temp 401C Man.temp Manual temperature value in case Temperature measurement is disabled R/W Float 2052
ma_fixed 401D Hold fix mA-output value when Hold-Fixed selected R/W Float 64
count_maint_time 401E M.due in Count-down maintenance value. Error 16 will be triggered when the count-down timer reaches 0 R Float 2052
itp_value 401F ITP Iso Thermal Point of intersection. Value in pH at which the temp. response of the system is at a null point R/W Float 12
reload_maint_time 4020 M.time Call for maintenance is a trigger to signal that the system has been in service for longer than the set time without calibration. The user can set up to 250 days as a routine service interval R/W Float 2052
aspot_low_limit 4022 R/W Float 2049
aspot_high_limit 4024 R/W Float 2049
slope_low_limit 4025 Slope L Slope low limit value for calibration acceptance R/W Float 2052
Name ID Label Description Type Access Class