Name ID Label Description Type Access Class
passcode_serv 4067 Service Passcode for service access in SC202 Possible values: 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 R/W Float 256
cc_actual 4068 CC act Actual cell constant after calibration R Float 2050
e6lim 4069 E6 lim. Limit for E6 (Conductivity too low) R/W Float 2050
e5lim 406A E5 lim. Limit for E5 (Conductivity too high) R/W Float 2050
sc_temp_sensor 406B Temp.sens Selection of the temperature compensation sensor R/W Enum 1024
sc_auto_return 406C Auto.Ret When Auto-return is ON, the converter reverts to the measuring mode from anywhere in the configuration menu, when no button is pressed during the set time interval of 10 minutes R/W Enum 1024
sc_percent 401C Percent When Percent by weight is ON, the 0% and 100% points can be entered R/W Enum 1024
sc_ma_burn 406D Burn func. Diagnostic error messages can signal a problem by sending the output signals upscale or downscale (22mA or 3.9mA). The Pulse burnout setting gives a 22mA signal for the first 30 seconds of an alarm condition. After the pulse the signal returns to normal to activate a latching alarm R/W Enum 1024
sc_polcheck 406E Pol.check When Polarisation check is ON, error 1 will be displayed in case of polirisation R/W Enum 1024
sc_fourelec 406F Electrodes Selection of the number of electrodes in the sensor R/W Enum 1024