名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
passcode_serv | 4067 | Service | Passcode for service access in SC202 Possible values: 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 | 读/写 | 浮点 | 256 |
cc_actual | 4068 | CC act | Actual cell constant after calibration | 读 | 浮点 | 2050 |
e6lim | 4069 | E6 lim. | Limit for E6 (Conductivity too low) | 读/写 | 浮点 | 2050 |
e5lim | 406A | E5 lim. | Limit for E5 (Conductivity too high) | 读/写 | 浮点 | 2050 |
sc_temp_sensor | 406B | Temp.sens | Selection of the temperature compensation sensor | 读/写 | 枚举 | 1024 |
sc_auto_return | 406C | Auto.Ret | When Auto-return is ON, the converter reverts to the measuring mode from anywhere in the configuration menu, when no button is pressed during the set time interval of 10 minutes | 读/写 | 枚举 | 1024 |
sc_percent | 401C | Percent | When Percent by weight is ON, the 0% and 100% points can be entered | 读/写 | 枚举 | 1024 |
sc_ma_burn | 406D | Burn func. | Diagnostic error messages can signal a problem by sending the output signals upscale or downscale (22mA or 3.9mA). The Pulse burnout setting gives a 22mA signal for the first 30 seconds of an alarm condition. After the pulse the signal returns to normal to activate a latching alarm | 读/写 | 枚举 | 1024 |
sc_polcheck | 406E | Pol.check | When Polarisation check is ON, error 1 will be displayed in case of polirisation | 读/写 | 枚举 | 1024 |
sc_fourelec | 406F | Electrodes | Selection of the number of electrodes in the sensor | 读/写 | 枚举 | 1024 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |