Method Name: meth_Calibrate
Method ID: 0x4254
Method Label: Calibrate
Method Help: Calibrate With this menu item you enable the calibration data determined during the measurements. The calibration data are transferred to "Meas Parameter". Thus, the detector will get a new calibration which in the future will be used to determine the measured values. After the calibration, a status message is displayed, indicating if the activation of the calibration data has been carried out successfully. If not, the measurement parameters are unchanged.
- {
- char status[3];
- SET_NUMBER_OF_RETRIES(7); _set_no_device(2); _set_all_resp_code(0); _set_device_status(0xFF,0); _set_xmtr_all_resp_code(0); _set_xmtr_device_status(0xFF,0);;
- ACKNOWLEDGE("|en|Attention: This will overwrite your measurement settings!");
- DELAY(0,"|en|Performing calibration ... 0%");
- send(240, status); if(status[0] != 0) { display_response_status(240, status[0]); process_abort(); };
- _iassign(16822,(3));
- DELAY(0,"|en|Performing calibration ... 15%");
- send(243, status);
- if((status[0] != 0) && (status[0] != 8))
- {