Loading…

Variable Name: var_dd_uiInstructDescriptor

Variable ID: 0x4021

Variable Label: Instruction Code

Variable Help:

Variable Handling: R

Variable Class: Dynamic | Diagnostic

Variable Size: 4

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 No instruction needed!
0x0001 1.Change - Electronics 2.Contact BT service
0x0002 1.Restart device 2.Change - Electronics 3.Contact BT service
0x0003 1.Restart device 2.Check - PMT connector - PMT detector unit - Electronics 3.Contact BT service
0x0004 1.Restart device 2.Check - EMC environment 3.Contact BT service
0x0005 1.Check - Detector temperature environment - Temperature alarm setup
0x0006 1.Check - Process - Process alarm setup
0x0007 1.Check - Detector settings - PMT connector - PMT detector unit - Electronics 2.Restart 3.Contact BT service
0x0008 1.Acknowledge event 2.Retry download 3.Try another file
0x0009 1.Retry download 2.Use calibration wizard for new configuration 3.Upload new configuration
0x000A 1.Acknowledge event 2.Restart 3.Check - Try to set Date/Time 4. Contact BT service
0x000B 1.Wait download... it will be finished automatically.
0x000C 1.Wait sampling... it will be finished automatically.
0x000D 1.Wait plateau measurement... it will be finished automatically.
0x000E 1.Wait calibration... it will be finished automatically.
0x000F 1.Exit simulation mode
0x0010 1.Check - Detector calibration - Pulse rate - Range values - Influence of radiation interference 2.Contact BT service
0x0011 1.Check - Detector calibration 2.Contact BT service
0x0012 1.This is a notifcation to check the activated alarm setup after a new calibration has been performed 1.Check device alarm setup 2.Acknowledge event
0x0013 1.Calibrate device by using the Quick Start Wizard
0x0014 1.Acknowledge event 2.Check actual measurement configuration 3.Run new calibration
0x0015 1.Acknowledge event 2.Run new calibration 3.Contact BT service
0x0016
0x0017
0x0018
0x0019
0x001A
0x001B 1.Reboot device 2.Change electronics
0x001C 1.Check - Configuration 2.Restart 3.Change electronics
0x001D 1.Restart device 2.Contact BT service
0x001E 1.Run factory reset 2.Run reboot
0x001F

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
143 4409 cmd_rd_error_info

Variable Write Command List

None