Loading…
Loading…
Variable Name: device_specific_status_15
Variable ID: 0x1032
Variable Label: Errors 308..315
Variable Help: Errors 308..315 Detector Control Errors Group 308..315
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0001 | 308 No impulses in control channel |
0x0002 | 309 No impulses in auxiliary channel |
0x0004 | 310 Impulse difference (measurement channel) |
0x0008 | 311 Impulse difference (control channel) |
0x0010 | 312 Impulse difference (auxiliary channel) |
0x0020 | 313 Instable pulse rates in both channels |
0x0040 | 314 Threshold of measurement channel 1 |
0x0080 | 315 Threshold of measurement channel 2 |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None