Variable Name: var_ds_eCalibrationMethodEdit

Variable ID: 0x41B9

Variable Label: Calibration Method

Variable Help: Calibration Method 1-POINT LIN Use the 1-point calibration only if full calibration is not possible. In this method, an empty adjustment (level below the monitoring level) is required. Using additional inputs such as product density and measuring path the count rate for the full calibration (level above the monitoring level) is then determined automatically. 2-POINT EXP An empty calibration (level below the monitoring level) and a full calibration (level above the monitoring level) is required for a two-point calibration.

Variable Handling: R/W

Variable Class: Correction

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 1-POINT
0x0001 2-POINT

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
239 423D cmd_rd_calibration_settings

Variable Write Command List

Command Number Command ID Command Name
240 423E cmd_wr_calibration_settings