Variable Name: var_ds_fHVDefault
Variable ID: 0x402E
Variable Label: HV Default
Variable Help: HV Default This setting affects: - A quick operation point control after power failure After a power failure, the HV starts at the last "HV Average" value which results from the "HV Default". This will significantly reduce the start-up phase after power failure. - Error detection in case of HV-drift If the HV deviates too far (+40%, -20%) from the HV Default or if the HV reaches the limits of the range of 300 V or 1300 V, this is reported as an error. - Special feature: 0V For maintenance purposes, the HV Default can also be set to 0V. In this case, the mechanisms described above are suspended; after a power failure, the measurement begins to get adjusted at 800 V. "HV Default" has already determined and set by BERTHOLD TECHNOLOGIES.
Variable Handling: R/W
Variable Class: Device
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|.0f
Variable Edit Format: |en|.0f
Variable Unit:
Variable Max:
Variable Min:
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
163 | 4205 | cmd_rd_detector_setup |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
164 | 4206 | cmd_wr_detector_startup |