变量名称: var_ds_fHVDefault

变量ID: 0x402E

变量标签: HV Default

变量描述: HV Default This setting affects: - A quick operation point control after power failure After a power failure, the HV starts at the last "HV Average" value which results from the "HV Default". This will significantly reduce the start-up phase after power failure. - Error detection in case of HV-drift If the HV deviates too far (+40%, -20%) from the HV Default or if the HV reaches the limits of the range of 300 V or 1300 V, this is reported as an error. - Special feature: 0V For maintenance purposes, the HV Default can also be set to 0V. In this case, the mechanisms described above are suspended; after a power failure, the measurement begins to get adjusted at 800 V. "HV Default" has already determined and set by BERTHOLD TECHNOLOGIES.

变量处理: 读/写

变量分类: 设备

变量长度: 4

变量数据类型: 浮点

变量显示格式: |en|.0f

变量编辑格式: |en|.0f

变量单位:

变量最大值:

变量最小值:

变量读取命令列表

命令号 命令ID 命令名称
163 4205 cmd_rd_detector_setup

变量写入命令列表

命令号 命令ID 命令名称
164 4206 cmd_wr_detector_startup