Loading…
Loading…
变量名称: var_ds_fUpperCalPoint
变量ID: 0x41C1
变量标签: Upper Cal Pt -Full
变量描述: Upper Cal Pt -Full Count rate with full container or at a level above the switching threshold. If you have selected in "Calibration Method" the setting 1-POINT, the value for "Upper Cal Pt-Full" is automatically calculated from the value of "Lower Cal Pt-Empty" and the parameters defined in the "Product Conditions menu". Select "ReadIn" to read-in the count rate, if you have chosen the setting "2-POINT" at Calibration Method. If you already know the value, you can enter it here.
变量处理: 读/写
变量分类: Correction
变量长度: 4
变量数据类型: 浮点
变量显示格式: |en|.0f
变量编辑格式: |en|.0f
变量单位:
变量最大值:
变量最小值:
变量读取命令列表
命令号 | 命令ID | 命令名称 |
---|---|---|
230 | 4239 | cmd_rd_upper_point |
变量写入命令列表
命令号 | 命令ID | 命令名称 |
---|---|---|
231 | 423A | cmd_wr_calibration_upper_point |