Loading…
Loading…
Variable Name: sc_caf
Variable ID: 0x4095
Variable Label: Calibration factor
Variable Help: This calibration factor is used to calculate the flow rate e.g. in kg/s for the specific sensor, the unit is in ( micro sec / kg * sec ). This factor is sensor specific and calculated during manufactories wet calibration and is stored in the SENSORPROM and stated on the sensor label.
Variable Handling: R/W
Variable Class: Device
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|5.6f
Variable Edit Format:
Variable Unit:
Variable Max: 99999.99
Variable Min: 99999.99
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
134 | 40F4 | read_float32 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
140 | 40F5 | write_float32 |