Variable Name: sc_cof

Variable ID: 0x4096

Variable Label: Correction factor

Variable Help: With this factor you are able to make a flow correction in % on the flowmeter. From factory the correction factor is set to 1.0000. Flow rate (kg/h) = Correction factor X flow rate (kg/h). If you want to change the flow on the flowmeter with +0,5 %, you have to change the correction factor to 1.005 in the device menu 'Sensor Characteristics'. After the change the flowmeter will now show a flow rate 0,5 % higher as before, on all flow - related values.

Variable Handling: R/W

Variable Class: Device

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|2.6f

Variable Edit Format:

Variable Unit:

Variable Max: 100

Variable Min: 100

Variable Read Command List

Command Number Command ID Command Name
134 40F4 read_float32

Variable Write Command List

Command Number Command ID Command Name
140 40F5 write_float32