Variable Name: sc_cof
Variable ID: 0x4096
Variable Label: Correction factor
Variable Help: With this factor you are able to make a flow correction in % on the flowmeter. From factory the correction factor is set to 1.0000. Flow rate (kg/h) = Correction factor X flow rate (kg/h). If you want to change the flow on the flowmeter with +0,5 %, you have to change the correction factor to 1.005 in the device menu 'Sensor Characteristics'. After the change the flowmeter will now show a flow rate 0,5 % higher as before, on all flow - related values.
Variable Handling: R/W
Variable Class: Device
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|2.6f
Variable Edit Format:
Variable Unit:
Variable Max: 100
Variable Min: 100
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
134 | 40F4 | read_float32 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
140 | 40F5 | write_float32 |