0 |
Unsigned8 |
1 |
User lock |
2 |
Servicemode |
3 |
Digital output |
4 |
Forced on |
5 |
Force output |
6 |
.1f |
7 |
Pulse output |
8 |
Pulse polarity |
9 |
Positive |
10 |
Negative |
11 |
Relay output |
12 |
Forced off |
13 |
.4f |
14 |
4.2f |
15 |
.2f |
16 |
mA |
17 |
Flowrate % |
18 |
Display the actual % of the 'Upper Value' (Qmax). |
19 |
.6f |
20 |
6.6f |
21 |
Mass flow |
22 |
No help available |
23 |
Volume flow |
24 |
Sensor temp. |
25 |
Density |
26 |
Flow direction |
27 |
Selection of the main flow direction. |
28 |
Mass flow Unit |
29 |
4.6f |
30 |
BS MFM Default |
31 |
Volume flow Unit |
32 |
bbl(31)/s |
33 |
Barrel per second; 1 bbl = 31 US gallons |
34 |
bbl(31)/min |
35 |
Barrel per minute; 1 bbl = 31 US gallons |
36 |
bbl(31) |
37 |
Barrel per hour; 1 bbl = 31 US gallons |
38 |
BS VFM Default |
39 |
BS VFM Max |
40 |
Density Unit |
41 |
t/m3 |
42 |
BS DMax Max |
43 |
BS DMax Min |
44 |
BS STMax Max |
45 |
BS STMax Min |
46 |
A |
47 |
Fraction A Help |
48 |
B |
49 |
Fraction B Help |
50 |
Pct. fract. A min. |
51 |
The measured procentage of media A in the fluid.Fraction is defined as a part of mixture. |
52 |
% |
53 |
Pct. fract. A max. |
54 |
Totalizer 1 unit |
55 |
Selects the unit of the totalizer 1. This unit is not updated in the keypad display unit of the device, only in the communication module of the device. To change the keypad display unit, use the parameter located in the 'Local Display' group. |
56 |
Barrel; 1 bbl = 31 US gallons |
57 |
Totalizer 2 unit |
58 |
Selects the unit of the totalizer 2.
This unit is not updated in the keypad display unit of the device, only in the communication module of the device.
To change the keypad display unit, use the parameter located in the 'Local Display' group. |
59 |
Total 1 function |
60 |
Select the measurement function |
61 |
Fraction A |
62 |
Fraction B |
63 |
Total 2 function |
64 |
1 |
65 |
Total flow. Totalizer 1 and Totalizer 2 can be reset individually. |
66 |
2 |
67 |
Totalizer 1 dir. |
68 |
Selection of the direction of flow for Totalizer 1 :
* Forward: Totalizer 1 only counts in forward direction of the measured flow.
* Reverse: Totalizer 1 only counts in backwards direction of the measured flow.
* Netto: Totalizer 1 counts in both directions of the measured netflow. |
69 |
Forward |
70 |
Reverse |
71 |
net |
72 |
Totalizer 2 dir. |
73 |
Selection of the direction of flow for Totalizer 1 :
* Forward: Totalizer 2 only counts in forward direction of the measured flow.
* Reverse: Totalizer 2 only counts in backwards direction of the measured flow.
* Netto: Totalizer 2 counts in both directions of the measured netflow. |
74 |
Low flow cut-off |
75 |
Here you select the cut-off between 0 and 10% of the maximum setting of the flow, equals the % value of the upper value.
This parameter is used to filter noise in the installation. It influences the display and all outputs, because it is forced to zero, if the flow is below this limit. |
76 |
Noise filter |
77 |
The MASS 6000 carries out signal processing using digital filtering. This technology allows sensor signals encumbered with noise to be filtered with different intensities. This applies to the mass flow measurement only. Setting 5 represents the maximum possible filtration and setting 1 represents the minimum possible filtration. Low filtering (setting 1) gives fast response to changes in flow and low immunity to noise. High filtering (setting 5) gives slow response to changes in flow and high noise immunity. |
78 |
1 * min |
79 |
2 ** |
80 |
3 *** |
81 |
4 **** |
82 |
5 ***** max |
83 |
Empty pipe |
84 |
* On: The error message for empty pipe is set if the measured density value is lower than the Empty pipe limit. * Off: The error message for empty pipe is never set. |
85 |
Empty pipe limit |
86 |
If the measured density is lower than this limit, the pipe is said to be empty and a warning/information message issued. |
87 |
g/Cucm |
88 |
Error level |
89 |
Sets the error level, which the flow meter will detect as an error.
* Fatal: Fatal errors are registered as errors
* Permanent: Fatal errors and permanent errors are registered as errors.
* Warning: Fatal errors, permanent errors and error warnings are registered as errors. |
90 |
Fatal |
91 |
Permanent |
92 |
Warning |
93 |
CO function |
94 |
Selects the measurement functionIt is not recommended to turn off the current output when communicating HART as it does not work with some masters. |
95 |
Pct. fraction A |
96 |
CO direction |
97 |
If unidirectional, only current output if flow is positive. If bidirectional, then current output in both direction. |
98 |
Unidirectional |
99 |
Bidirection |
100 |
CO range |
101 |
* 0-20 mA: This setting is in case of HART communication not possible.
* 4-20 mA: Used for HART communication without alarm output.
* 4-20 mA + Alarm: The current output gives following alarm signals:
* Fatal: 1.3 mA
* Permanent: 2 mA
* Warning Error: 3 mA
The error level, which the flow meter will detect as error, can be set in Error Level parameter in the Diagnostic group. |
102 |
4-20 mA |
103 |
4-20 mA+Alarm |
104 |
CO timeconstant |
105 |
Use to filter the output when having noise. This setting also influence the display and relay. |
106 |
DO select: |
107 |
Selection of digital output from the list:
* Pulse: Output is active as pulse output.
* Frequency: Output is active as frequency output.
* Error level: Output is 'High' when error occurs.
Error level is selected by Error Level parameter in Diagnostics group.
* Error number: Output is 'High', when selected error occurs.
* Direction/limit: Output is 'High', when selected flow value is achieved.
* Batch: Digital Output is used for batch control:
Batching a preset quantity is controlled by the flow meter. The output can be used to control i.e. a valve; the output is normally low and the valve is closed. Starting the batching process (from Digital Input, Keypads display or PDM) sets the output 'High' until the preset Batch quantity is reached.
The digital- and relay-output cannot be set to batch at the same time. If the digital output is already used as batch output and the external input as batch input, the relay output will be set to OFF. |
108 |
Pulse |
109 |
Frequency |
110 |
Error number |
111 |
Direction/limit |
112 |
Batch |
113 |
Pulse function |
114 |
Mass/pulse |
115 |
Specifies the mass amount for each pulse. |
116 |
.9f |
117 |
6.9f |
118 |
Os Do1 Mp Default |
119 |
Os Do1 Mp Min |
120 |
Quadrature |
121 |
This is not recommended in a HART application. Quadrature set ON will set current output OFF. |
122 |
Volume/pulse |
123 |
Specifies the volume amount for each pulse. |
124 |
Cum |
125 |
Os Do1 Vp Default |
126 |
Os Do1 Vp Min |
127 |
Pulse flow dir. |
128 |
Pulse width |
129 |
Pulse width of Digital Output1. |
130 |
64 µs |
131 |
130 µs |
132 |
260 µs |
133 |
510 µs |
134 |
1.0 ms |
135 |
2.0 ms |
136 |
4.1 ms |
137 |
8.2 ms |
138 |
16 ms |
139 |
33 ms |
140 |
66 ms |
141 |
130 ms |
142 |
260 ms |
143 |
520 ms |
144 |
1.0 s |
145 |
2.1 s |
146 |
4.2 s |
147 |
Digital output pulse polarity. Can be set according to the specification of the pulse counter. |
148 |
Frequency function |
149 |
Frequency flow dir. |
150 |
Fmax output |
151 |
Maximum frequency |
152 |
10 kHz |
153 |
5 kHz |
154 |
1 kHz |
155 |
500 Hz |
156 |
Freq. time const. |
157 |
Use to filter the output when having noise. Normally allways set to 0.1. Only to be changed in very noisy enviroment. |
158 |
DO limit function |
159 |
DO limit mode |
160 |
Selects one or two setpoints. |
161 |
1 setpoint |
162 |
2 setpoints |
163 |
DO limit setp. 1 |
164 |
DO limit setpoint |
165 |
Used only in case of two setpoints: Selection of the minimum % value (of the Upper Value of the range), above which the output shall be active. |
166 |
To select the % of Upper Value, where the output shall be active.
If 'Direction/limit' is selected for the Relay output, set this value to 0.00% |
167 |
DO limit setp. 2 |
168 |
Used only in case of two setpoints: Selection of the maximum % value (of the Upper Value of the range), above which the output shall be inactive. |
169 |
DO limit hyst. |
170 |
To select % of Upper Value. This function is used to make a difference when the setpoint is switching from active to not active. When 1 setpoint is used for direction mode, Hysteresis should be set to 5 %.Hysteresis in % of the Upper Value: The relay will be active when the measurement is greater than the upper limit. The relay contains beeing active, until the output limis goes beyond the Alarm Limit minus hysteresis. |
171 |
To select an error number on which the output will be active 'on'. |
172 |
Batch function |
173 |
Batch quantity |
174 |
Amount to be batched.
The batch quantity cannot be less than the batch compensation value. |
175 |
Amount |
176 |
The actual batch value, which has been reached during the batching process. |
177 |
Batch direction |
178 |
* Upwards: Batch starts to count from zero and up to batch quantity. * Downwards: Batch starts to count from batch quantity and down to zero. |
179 |
Up |
180 |
Down |
181 |
Batch unit |
182 |
This parameter defines the unit for the batch output.This unit is not updated in the keypad display unit of the device, only in the communication module of the device. To change the keypad display unit, use the parameter located in the 'Local Display' group. |
183 |
Batch compensation |
184 |
Offset value to compensate for application delays. The maximum batch compensation cannot be set more than the batch quantity value. The batch quantity value has to be increased before batch compensation is increased. |
185 |
Batch lead const. |
186 |
Time constant is used to compensate the batch quantity.
This compensation takes into account flow variations.
How to set it:
Start by setting all compensation to zero (Compensation and lead constant). Enter desired amount, M-wanted (ex. 25kg) and start batch. Observe the flow reading, Q-end (ex. 5000kg/h) just before the batch stops. Observe, what amount the meter totalizes (in the display) M-real (ex. 25.5kg).
Calculate the lead constant as:
LeadConst = (M-real - M-wanted) - Q-end
Remember to get the correct units; LeadConst is in seconds.
In our example: LeadConst = (25.5kg-25kg)/(5000kg/h) = 0.0001h x 3600sec/h = 0.36 seconds
Enter this value as lead constant. This is a good starting point, only small adjustments can be necessary. Mass 6000 should now fill 25 kg, also if the flow changes (if the valve closing time also changes, this will not be the case! - The lead constant can only compensate for true flow changes).
Remember to set the lead constant back to zero before leaving the batch menu, if using pulse or frequency output instead! Otherwise the output is still being compensated. |
187 |
|
188 |
Batch time error |
189 |
Specification of batch time error:
* On: Sets a maximum time for the batch. If batch quantity is not achieved within the max batch time, the error will be indicated.
* Off: No batch time error will be indicated. |
190 |
Max. batch time |
191 |
Sets a maximum time for the batch. If batch quantity is not achieved within the max batch time the error will be indicated. |
192 |
Batch overrun err. |
193 |
Specification of batch overrun error:
* On: If the amount exceeds the maximum amount for the batch an error will be indicated.
* Off: No batch overrun error will be indicated. |
194 |
Max. batch overrun |
195 |
Sets a maximum amount for the batch overrun. If amount exceeds the maximum an error will be indicated. |
196 |
Mode: |
197 |
Show the batch status (Batch active, Stopped, or Paused) |
198 |
Stopped |
199 |
Active |
200 |
Paused |
201 |
BATCH NOT ACTIVE! |
202 |
Cycle cnt. |
203 |
Number of started batches |
204 |
RLO select: |
205 |
Select relay output from the drop down list. |
206 |
RLO limit function |
207 |
RLO limit mode |
208 |
RLO limit setp. 1 |
209 |
RLO limit setpoint |
210 |
RLO limit setp. 2 |
211 |
RLO limit hyst. |
212 |
DI Select |
213 |
Selection of the external input from the list. The selected input will be active, when the external input (no. 77 and 78) is activated
* Start Batch: Activates the batching process.
* Hold/Continue: Batch will be paused while external input is activated.
* Stop Batch: Stops the batching process.
* Zero Adjust: Activates the Zero adjust process.
* Totalizer Reset: Selected totalizer will be reset. Either totalizer 1, totalizer 2 or both can be selected.
* Force output: Flow meter will be forced on x(selected) % of Upper Value (-125% to +125%).
* Freeze output: Flow meter will be frozen on the actual flow.
If the input is selected as Batch input, the digital output will be set to batch as default. |
214 |
Start batch |
215 |
Hold/continue batch |
216 |
Stop batch |
217 |
Zero adjustment |
218 |
Totalizer reset |
219 |
Freeze output |
220 |
This parameter will force the flow meter to % of Upper Value (-125% to +125%). |
221 |
.0f |
222 |
Selects the totalizer to be reset by the external input |
223 |
1 reset |
224 |
2 reset |
225 |
1+2 reset |
226 |
Eis Ei Tmp1 |
227 |
Eis Ei Tmp2 |
228 |
SENSORPROM: |
229 |
Indicates, if the Sensorprom is installed in the transmitter. If not, the sensor data must be programmed manually. |
230 |
Suppress P40 |
231 |
* Yes: The device will not indicate an error, when the sensorprom is missing.
* No: The device will indicate error, when the sensorprom is missing.
This parameter is always set to 'No' after power on. |
232 |
Sensor size |
233 |
Sensor Diameter Di in mm. This information is written on the sensor label. |
234 |
1.6f |
235 |
Calibration factor |
236 |
This calibration factor is used to calculate the flow rate e.g. in kg/s for the specific sensor, the unit is in ( micro sec / kg * sec ). This factor is sensor specific and calculated during manufactories wet calibration and is stored in the SENSORPROM and stated on the sensor label. |
237 |
5.6f |
238 |
µs²/kg |
239 |
Correction factor |
240 |
With this factor you are able to make a flow correction in % on the flowmeter. From factory the correction factor is set to 1.0000. Flow rate (kg/h) = Correction factor X flow rate (kg/h). If you want to change the flow on the flowmeter with +0,5 %, you have to change the correction factor to 1.005 in the device menu 'Sensor Characteristics'. After the change the flowmeter will now show a flow rate 0,5 % higher as before, on all flow - related values. |
241 |
2.6f |
242 |
Sensor TC |
243 |
Temperature coefficient 'Sensor TC'(%/DegC) for the specific sensor is found under the flow calibration.
On the MASS2100 sensor is mounted a PT1000 temperature sensor and on the MC1 a PT100.
With this temperature, with which the sensors are mounted, the transmitter is able to compensate for temperature changes in the sensors. |
244 |
%/DegC |
245 |
Density parm. A |
246 |
Density parameter A is a constant found under the density calibration.
This constant is used in the formula to calculate the density of the fluid in the sensor.
Density (kg/m3 ) = A + B (1 + Density TC x temp.) * (1/(fr)^2).
A = Density parameter A (kg/m3).
B = constant.
Density TC = Sensor`s density temperature coefficient ( %/DegC).
fr. = the sensors resonance freq(Hz).
Standard sensors without density calibration have average values stored in the sensorprom to be able to monitor the density. |
247 |
kg/Cum |
248 |
Density parm. B |
249 |
Density parameter B is a constant found under the density calibration.
This constant is used in the formula to calculate the density of the fluid in the sensor.
Density (kg/m3 ) = A + B (1 + Density TC x temp.) * (1/(fr)^2).
A = Density parameter A (kg/m3).
B = constant.
Density TC = Sensor`s density temperature coefficient ( %/DegC).
fr. = the sensors resonance freq(Hz).
Standard sensors without density calibration have average values stored in the sensorprom to be able to monitor the density. |
250 |
M |
251 |
Density TC |
252 |
Density TC the temperature coefficient ( %/DegC ) is found under the density calibration.
This constant is going into the formula to calculate the density of the fluid in the sensor.
Density ( kg/m3 ) = A + B ( 1 + Density TC x temp.) * (1/(fr)^2).
A = Density parameter A ( kg/m3 )
B = constant
Density TC = Sensor density temperature coefficient ( %/DegC)
fr. = Sensor resonance frequency ( Hz )
Standard sensors without density calibration have average values stored in the sensorprom to be able to monitor the density. |
253 |
Density offset |
254 |
With this function you are able to make an offset on the measured density.
From factory the density offset is set to 00,000 kg/m3.
If you want the flowmeter to show + 2 kg/m3 more, then you can change the density offset to 02,000 Kg/m3 in the 'Sensor' menu. |
255 |
Density factor |
256 |
With this factor you are able to make a density correction in % on the flowmeter.
From factory the correction factor is set to 1.0000.
Density ( kg/m3 ) = Correction factor x density ( kg/m3 ).
If you want to change the density on the flowmeter with +0,5 %, you have to change the correction factor to 1.005 in the device menu 'Sensor Characteristics'
After the change the flowmeter will now show a density about 0,5 % higher as before. |
257 |
Fraction offset |
258 |
Fraction is defined as a part of a mixture.
Mixtures consist of two components (A+B), which can be measured separately. If the flowmeter is ordered with fraction, e.g. °BRIX, the flowmeter is able to calculate the % concentration of sugar in a water (B) + sugar (A) mixture. With this fraction offset you are able to make an absolute offset on the measured concentration.
Formula:
% concentration = ( a / 3600 ) + bx
% concentration = the concentration of media e.g. in °BRIX
a = fraction offset in ( kg/h )
b = is a factor without dimension ( table slope )
x = is the concentration in % or e.g. °BRIX
Example:
If you want to make an offset on + 1 °BRIX you have to make following calculation:
a/3600 has to be 1%.
a/3600 =0,01
a = 0,01x3600 = 36 kg/h
Fraction A is in this example the calculated flow rate of 100 % pure sugar,
and Fraction B is the calculated flow rate of 100 % pure water. |
259 |
Table slope |
260 |
Fraction is defined as a part of mixture.
These mixtures consist of two components (A+B), which can be measured separately. If the flowmeter is ordered with fraction, e.g. °BRIX, it is able to calculate the % concentration of sugar in a water (B) + sugar (A) mixture. From factory the table slope 'b' is set to 1.0000.
Formula:
% concentration = ( a / 3600 ) + bx
% concentration = the concentration of media e.g. in °BRIX
a = fraction offset in ( kg/h )
b = is a factor without dimension ( table slope )
x = is the concentration in % or e.g. °BRIX
Example:
If you want to change the concentration in % on the flowmeter with +0,5 %, you have to change the table slope 'b' to 1.005.
Afterwards the flowmeter will show a concentration 0,5 % higher as before. |
261 |
Operating time |
262 |
Counts the days, during which the device has been in operation. |
263 |
5d |
264 |
Zero adjust |
265 |
Select between manual and auto zero adjustment |
266 |
Auto |
267 |
Manual |
268 |
Zero adjust state |
269 |
Sensor zero |
270 |
The sensor can be set manually or automatically during the Zero adjustment. |
271 |
rm zasz min |
272 |
rm zasz max |
273 |
Zero adjust time |
274 |
Sets the zero adjust time: The time it takes to perform a zero adjustment |
275 |
zero adjust error state |
276 |
Zero sigma |
277 |
set the zero sigma: |
278 |
Sigma limit |
279 |
Sets the limit for the sigma value. |
280 |
Transmitter type |
281 |
The type of the Transmitter (e.g. MASS6000). |
282 |
Type |
283 |
Transmitter MLFB |
284 |
This is the order number (MLFB) of the product.
This information is written on the product label. |
285 |
Transm. serial no |
286 |
This number is the serial number of the product . When the MFLB number and the serial number are known, it is possible to find the data for the specific product e.g. the calibration data. |
287 |
Transmitter FW ver. |
288 |
Firmware version of the Transmitter |
289 |
Sensor MLFB no |
290 |
This is the MLFB number, ordering -/ article number of the product. This information is written on the sensor label. |
291 |
Sensor serial no |
292 |
This number is the serial number of the product.
When the MLFB number and the serial number are known, it is possible to find the data for the specific product e.g. the calibration data. |
293 |
Customer code no |
294 |
Text string that can be used for additional information. |
295 |
HART FW version |
296 |
Firmware version of the HART module (communication module) |
297 |
HART HW version |
298 |
Hardware version of the HART module (communication module) |
299 |
CANAPP |
300 |
attr |
301 |
Qualifier |
302 |
Save parameter |
303 |
data unit |
304 |
string length |
305 |
Octet string byte 1 |
306 |
Octet string byte 2 |
307 |
LD language |
308 |
All the text on the keypad display appears in the chosen language. |
309 |
English |
310 |
German |
311 |
French |
312 |
Danish |
313 |
Swedish |
314 |
Finnish |
315 |
Russian |
316 |
Spanish |
317 |
Italian |
318 |
Portuguese |
319 |
Write Select |
320 |
Sensor size (text) |
321 |
Sensor Size Mass Help |
322 |
Sensor type |
323 |
Sigma limit min |
324 |
Sigma limit min Help |
325 |
Sigma limit max |
326 |
Sigma limit max Help |
327 |
LD sens temp. unit |
328 |
Units Dislay Help |
329 |
LD density unit |
330 |
kg/Cuft |
331 |
kg/Cuin |
332 |
kg/Cucm |
333 |
ton/Cum |
334 |
ton/Cuft |
335 |
ton/Cuin |
336 |
ton/Cucm |
337 |
lb/Cum |
338 |
lb/Cuft |
339 |
lb/Cuin |
340 |
lb/Cucm |
341 |
mg/Cum |
342 |
mg/Cuft |
343 |
mg/Cuin |
344 |
mg/Cucm |
345 |
g/Cum |
346 |
g/Cuft |
347 |
g/Cuin |
348 |
LD pulse Unit |
349 |
kg |
350 |
ton |
351 |
lb |
352 |
mg |
353 |
g |
354 |
ml |
355 |
l |
356 |
hl |
357 |
kl |
358 |
Ml |
359 |
US BBL |
360 |
Cuft |
361 |
Cuin |
362 |
US G |
363 |
US kG |
364 |
US MG |
365 |
UK G |
366 |
UK kG |
367 |
UK MG |
368 |
LD batch unit |
369 |
LD total. 2 unit |
370 |
LD total. 1 unit |
371 |
LD vol. flow unit |
372 |
Volumeflow Units Dislay Help |
373 |
Cum/s |
374 |
Cum/min |
375 |
Cum/h |
376 |
Cum/d |
377 |
ml/s |
378 |
ml/min |
379 |
ml/h |
380 |
ml/d |
381 |
l/s |
382 |
l/min |
383 |
l/h |
384 |
l/d |
385 |
hl/s |
386 |
hl/min |
387 |
hl/h |
388 |
hl/d |
389 |
kl/s |
390 |
kl/min |
391 |
kl/h |
392 |
kl/d |
393 |
Ml/s |
394 |
Ml/min |
395 |
Ml/h |
396 |
Ml/d |
397 |
US BBL P S |
398 |
US BBL P M |
399 |
US BBL P H |
400 |
US BBL P D |
401 |
Cuft/s |
402 |
Cuft/min |
403 |
Cuft/h |
404 |
Cuft/d |
405 |
Cuin/s |
406 |
Cuin/min |
407 |
Cuin/h |
408 |
Cuin/d |
409 |
US G P S |
410 |
US G P M |
411 |
US G P H |
412 |
US G P D |
413 |
US MG P S |
414 |
US MG P M |
415 |
US MG P H |
416 |
US MG P D |
417 |
UK G P S |
418 |
UK G P M |
419 |
UK G P H |
420 |
UK G P D |
421 |
UK MG P S |
422 |
UK MG P M |
423 |
UK MG P H |
424 |
UK MG P D |
425 |
US kG P S |
426 |
US kG P M |
427 |
US kG P H |
428 |
US kG P D |
429 |
UK kG P S |
430 |
LD mass flow unit |
431 |
kg/s |
432 |
kg/min |
433 |
kg/h |
434 |
kg/d |
435 |
ton/s |
436 |
ton/min |
437 |
ton/h |
438 |
ton/d |
439 |
lb/s |
440 |
lb/min |
441 |
lb/h |
442 |
lb/d |
443 |
mg/s |
444 |
mg/min |
445 |
mg/h |
446 |
mg/d |
447 |
g/s |
448 |
g/min |
449 |
g/h |
450 |
g/d |
451 |
Fraction A max. |
452 |
Defines the maximum value of fraction A in flow unit. |
453 |
Fraction B max. |
454 |
Defines the maximum value of fraction B in flow unit. |
455 |
Totalizer 2.. |
456 |
Totalizer 2 is controlled by batch. |
457 |
Driver signal |
458 |
Sensor 1 amplitude |
459 |
mV |
460 |
Sensor 2 amplitude |
461 |
Sensor frequency |
462 |
Hz |
463 |
Fraction Unit |
464 |
pi_swv_integer16 |
465 |
sm_rlo_pol |
466 |
Positive |
467 |
Negative |
468 |
Pct. A. |
469 |
The measured percentage of media A in the fluid. Fraction is defined as a part of mixture.
Mixtures consisting of two components ( A+ B) which can be measured separately.If the flowmeter is ordered with fraction, e.g °BRIX,the flowmeter is able to calculate the % concentration of sugar in a water (B) + sugar (A) mixture.From factory the table slope 'b' is set to 1.0000.
Formula:
% concentration = (a / 3600) + bx
% concentration = the concentration of media e.g. in °BRIX
a = fraction offset in (kg/h)
b = is a factor without dimension (table slope)
x = is the concentration in % or e.g. °BRIX
Example:
If you want to change the concentration in % on the flowmeter with +0,5 %, you have to change the table slope 'b' to 1.005.
After the change the flowmeter will now show a concentration 0,5 % higher as before. |
470 |
set |
471 |
check_valid_mfm_unit |
472 |
check_valid_vfm_unit |
473 |
check_valid_t2_unit_m |
474 |
check_valid_t2_unit_v |
475 |
check_valid_t1_unit_m |
476 |
check_valid_t1_unit_v |
477 |
check_valid_bat_unit_m |
478 |
check_valid_bat_unit_v |
479 |
EI Selection Adjustment |
480 |
Digital Input Reset Digital output |
481 |
Relay output Reset Digital output |
482 |
Digital output Reset Relay output |
483 |
Digital output pulse pre write |
484 |
Frequensy output pulse pre write |
485 |
os_do1_lim_select_pre_write |
486 |
os_rlo1_lim_select_pre_write |
487 |
Totalizer 1 reset |
488 |
Resets the totalizer 1 |
489 |
Totalizer 2 reset |
490 |
Resets the totalizer 2 |
491 |
Batch start |
492 |
Activates the batching process. |
493 |
Batch pause |
494 |
Pauses the batching process. |
495 |
Batch stop |
496 |
Stops the batching process. |
497 |
Batch resume |
498 |
Resumes the batching process. |
499 |
Reset cycle cnt. |
500 |
Resets the cycle counter. |
501 |
check_valid_ffm_unit |
502 |
Relay output simulation |
503 |
Digital output simulation |
504 |
Zero adjust manual |
505 |
Zero adjust auto |
506 |
Reset to default |
507 |
Exit service mode |
508 |
Release user lock |
509 |
DD Version |
510 |
Version number of the device description according to NAMUR specification. |
511 |
Select current output |
512 |
Device setup.. |
513 |
Identification.. |
514 |
Device.. |
515 |
Sensor.. |
516 |
Transmitter.. |
517 |
Input.. |
518 |
Digital input DI.. |
519 |
Output.. |
520 |
Process variables.. |
521 |
Mass flow.. |
522 |
Volume flow.. |
523 |
Density.. |
524 |
Sensor temp.. |
525 |
Fraction.. |
526 |
Totalizer 1.. |
527 |
Current output CO.. |
528 |
Digital output DO.. |
529 |
Relay output RLO.. |
530 |
Diagnostic.. |
531 |
Local display LD.. |
532 |
Operation.. |
533 |
Totalizer.. |
534 |
Simulation.. |
535 |
Batch.. |
536 |
Batch setup.. |
537 |
Hot key |
538 |
SITRANS FC MASSFLO |
539 |
SIEMENS |
540 |
Warning Errors |
541 |
Permanent Errors |
542 |
Fatal Errors |
543 |
Reserved |
544 |
Field Device Status 1 Help |
545 |
Fatal F60 CAN comm. error Converter/AOM |
546 |
Fatal F61 SENSORPROM err. Replace |
547 |
Fatal F62 SENSORPROM ID Replace |
548 |
Fatal F63 SENSORPROM Replace |
549 |
Fatal F70 Coil current Check cables |
550 |
Fatal F71 Driver phase Check cables/pol. |
551 |
Fatal F80 Internal error Restart or replace |
552 |
Fatal F81 Internal error Restart or replace |
553 |
Field Device Status 2 Help |
554 |
Fatal F82 Internal error Restart or replace |
555 |
Fatal F83 Internal error Restart or replace |
556 |
Fatal F84 Sensor level Sensor jammed |
557 |
Fatal F96 Inst. module Restart or replace |
558 |
Fatal F97 AOM to old Replace |
559 |
F64 CONV.PROM err.,Replace |
560 |
Field Device Status 3 Help |
561 |
Field Device Status 4 Help |
562 |
Field Device Status 5 Help |
563 |
Field Device Status 14 Help |
564 |
Permanent P40 SENSORPROM Insert/change |
565 |
Permanent P41 Parameter range Switch off and on |
566 |
Permanent P42 Current output Check cables |
567 |
Permanent P43 Internal error Switch off and on |
568 |
Permanent P44 CT SENSORPROM Replace |
569 |
Field Device Status 15 Help |
570 |
Permanent P49 Protect. viol. Switch off and on |
571 |
Permanent P50 Temp. cable Check cables. |
572 |
Permanent P51 ASIC overflow Check installation |
573 |
Permanent P52 ASIC overflow Check installation |
574 |
Permanent P53 ASIC overflow Check installation |
575 |
Field Device Status 16 Help |
576 |
Permanent P54 ASIC overflow Check installation |
577 |
Permanent P55 ASIC overflow Check installation |
578 |
Permanent P56 ASIC overflow Check installation |
579 |
Permanent P57 ASIC overflow Check density |
580 |
Permanent P58 ASIC overflow Check installation |
581 |
Field Device Status 17 Help |
582 |
Field Device Status 18 Help |
583 |
Warning W20 Totalizer Reset manually |
584 |
Warning W21 Pulse overflow Adj. pulse settings |
585 |
Warning W22 Batch timeout Check installation |
586 |
Warning W23 Batch overrun Check installation |
587 |
Warning W24 Batch neg. flow Check flowdirection |
588 |
Warning W30 Flow saturated Check installation |
589 |
Warning W31 Empty pipe Check density |
590 |
Field Device Status 19 Help |
591 |
Field Device Status 20 Help |
592 |
Warning W32 Temp. too high Adjust temperature |
593 |
Warning W33 Temp. too low Adjust temperature |
594 |
Warning W34 Zero adj. fail Check flow = zero |
595 |
Warning W35 Current out 1 Check max. settings |
596 |
Warning W36 Frequency out 1 Check max. settings |
597 |
Field Device Status 21 Help |
598 |
Field Device Status 22 Help |
599 |
Field Device Status 23 Help |
600 |
Field Device Status 24 Help |
601 |
unit |
602 |
FMU860 |
603 |
DT |
604 |
DVC |
605 |
FCX |
606 |
AS800 |
607 |
340T |
608 |
340C |
609 |
344 |
610 |
Siemens Micro K |
611 |
SITRANS L |
612 |
SIPAN PH |
613 |
SITRANS P |
614 |
75SMT |
615 |
UNIDELTA |
616 |
8320 |
617 |
SX5100 |
618 |
TSV175 |
619 |
DMU |
620 |
3780 |
621 |
BM70 |
622 |
8u |
623 |
This operation sets the relay output to fixed mode.
It does not correspond to the process value anymore. |
624 |
Please wait... |
625 |
Simulation relay is active and in normal mode - press OK to terminate. |
626 |
Simulation relay is active and forced to %{local_sm_rlo_rlofv} (Press OK to terminate) |
627 |
This operation sets the digital output to fixed mode. It does not correspond to the process value anymore. |
628 |
Digital output is set to frequency. |
629 |
Frequency value (min %{fMin} ... max %{fMax} Hz) |
630 |
The entered frequency value was invalid!Frequency (min %{fMin} ... max %{fMax} Hz) |
631 |
loc_sm_do_dofv |
632 |
Simulation digital output with frequency %{loc_sm_do_dofv} Hz is active (Press OK to terminate) |
633 |
Digital output is set to pulse. |
634 |
Forced value (High or Low) |
635 |
Simulation digital output pulse is active and forced to %{loc_sm_do_dopv} (Press OK to terminate) |
636 |
Simulation digital output pulse is active in normal mode - press OK to terminate. |
637 |
This operation sets the output current to fixed mode.It does not |
638 |
Current (%{fMin} .. 24.0 mA) |
639 |
fCurrentToSimulate |
640 |
The entered current value was invalid! Current (%{fMin} .. 23.0 mA) |
641 |
Simulation analog output %{fCurrentToSimulate} mA active (Press OK |
642 |
Zero adjust point has been set - press OK to continue. |
643 |
Do you wish to perform zero adjustment? |
644 |
Zero adjust in process
- Please wait... |
645 |
Zero adjust auto has finished successfully. |
646 |
Zero adjust auto has finished with error.
Sigma limit exceeded maximum. |
647 |
Zero adjust auto has finished with error.
Zero adjust exceeded maximum. |
648 |
Zero adjust auto has finished with error.
Zero adjust status could not be evaluated, try again. |
649 |
Reset to factory settings? - This will take approx. 30 sec. |
650 |
Setting factory defaults! - Please wait... |
651 |
Saving factory defaults!. - Please wait... |
652 |
The device has been reset successfully. |
653 |
DD Version
01.01.02 |
654 |
* Off: It is not recommended to turn Off the current output, please turn On. |