Loading…
Loading…
方法名称: meth_SetSafetyModeOn
方法ID: 0x4251
方法标签: Safety ON
方法描述: Safety ON The safety mode has to be enabled for safety-relevant applications. In safety mode, safety-limiting values are set to a safety-compatible setting. In addition, you will be prompted to protect the detector with a password. The following settings are set automatically when safety mode is enabled: Current Loop Monitoring: = ENABLED HV Mode = AUTO Test Settings = NORMAL Error Handling = SENSITIV Digital inputs: = OFF Multidrop: Poll Addr = 0
- {
- char status[3];
- int i;
- SET_NUMBER_OF_RETRIES(7); _set_no_device(2); _set_all_resp_code(0); _set_device_status(0xFF,0); _set_xmtr_all_resp_code(0); _set_xmtr_device_status(0xFF,0);;
- send(143, status); if(status[0] != 0) { display_response_status(143, status[0]); process_abort(); };
- if(_ivar_value(16398) == _ivar_value(16400))
- {
- ACKNOWLEDGE("|en|Safety is already activated!");
- }
- else
- {
- send(15, status); if(status[0] != 0) { display_response_status(15, status[0]); process_abort(); };
- if(_ivar_value(167) == _ivar_value(16384))
- {
- _get_dev_var_value(("|en|Enter Password:"),0,16388);
- _vassign(16398,16400);
- send(144, status); if(status[0] != 0) { display_response_status(144, status[0]); process_abort(); };