Variable Name: var_dd_eHVMode

Variable ID: 0x402C

Variable Label: HV Mode

Variable Help: HV Mode The HV-mode (high-voltage control) provides the following options: AUTO Automatic HV control. Select this mode in normal measurement mode. It enables a temperature-stable operation of the detector. MANUAL Manual HV control. This setting is used for testing purposes; you can also provisionally set the operating point of the detector. The specified voltage must, however, lie in the plateau. If you select "Manual", the automatic HV control is switched off. The HV is then set to the value that was set in the menu item HV Manual. PLATEAU Starts the plateau measurement. If you select PLATEAU, the measurement mode is exited and the plateau of the multiplier measured, which is used in the detector.

Variable Handling: R/W

Variable Class: Dynamic | Device

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 AUTO
0x0001 MANUAL
0x0002 PLATEAU

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
166 4208 cmd_rd_dynamic_probe_values

Variable Write Command List

Command Number Command ID Command Name
167 4209 cmd_wr_hv_mode