变量名称: var_dd_eHVMode
变量ID: 0x402C
变量标签: HV Mode
变量描述: HV Mode The HV-mode (high-voltage control) provides the following options: AUTO Automatic HV control. Select this mode in normal measurement mode. It enables a temperature-stable operation of the detector. MANUAL Manual HV control. This setting is used for testing purposes; you can also provisionally set the operating point of the detector. The specified voltage must, however, lie in the plateau. If you select "Manual", the automatic HV control is switched off. The HV is then set to the value that was set in the menu item HV Manual. PLATEAU Starts the plateau measurement. If you select PLATEAU, the measurement mode is exited and the plateau of the multiplier measured, which is used in the detector.
变量处理: 读/写
变量分类: 动态 | 设备
变量长度: 1
变量数据类型: 枚举
变量枚举值 | 变量枚举文本 |
---|---|
0x0000 | AUTO |
0x0001 | MANUAL |
0x0002 | PLATEAU |
变量显示格式:
变量编辑格式:
变量单位:
变量最大值: N/A
变量最小值: N/A
变量读取命令列表
命令号 | 命令ID | 命令名称 |
---|---|---|
166 | 4208 | cmd_rd_dynamic_probe_values |
变量写入命令列表
命令号 | 命令ID | 命令名称 |
---|---|---|
167 | 4209 | cmd_wr_hv_mode |